GE KrautKramer USN 60 Ultrasonic Flaw Detector Thickness Gauge
Original – GE KrautKramer USN 60 Ultrasonic Flaw Detector
GE Inspection Technologies has optimized its USN 60 series for use in direct sunlight and operation at extreme temperatures. These new characteristics make the instruments ideally suited for outdoor use with its increased long battery operation time.
The combination of the rugged USN durability, 11 hours of battery operation, fast rotary knob operation, outstanding ultrasonic performance, and the モsquare wave pulserヤ form a powerful portable ultrasonic inspection tool.
Optimized outdoor use design
Extreme temperature use (-20°C to +55°C /-4°F to 130°F)
Easy to view in direct sunlight
11 hours battery operation
Vibrant colors
Hi-resolution color LCD display produces モAnalog Look and Performanceヤ echo dynamics.
4 selectable vibrant display color schemes to match lighting conditions.
Gates and gate functions are color coded for easy identification and fast adjustment.
User Preferred Features
Simple operation with fast rotary knob adjustments; gain is always directly accessible with the left-hand rotary knob and lockable.
Auto CAL makes calibration fast & easy.
15 Hz to 6 kHz (spike mode) PRF and 15 Hz to 2 kHz (square mode) PRF (pulse repetition frequency)
2 independent gates monitor amplitude and soundpath distance for both flaw detection and thickness measurement applications.
250 KHz to 25 MHz frequency range.
RF display mode enhances signal evaluation and bond inspection of dissimilar materials.
4 selectable damping settings (50, 75, 150, 500 ohms) for optimum probe performance.
1 mm to 28 m (0.040ヤ to 1100ヤ) range (in steel) covers thin to lengthy acoustically clean materials.
dB REF key evaluates subsequent echoes gain value and amplitude against the highest echo in Gate A (reference echo) when activated.
IF (Interface) Gate Option for automatic start of the display, Gate A, Gate B, and / or DAC / TCG for immersion testing applications.
VGA Output Option provides an easy way to connect to a PC monitor or PC projector for viewing by large audiences or training purposes.
RF Output Option outputs the raw RF waveform via a standard Lemo connector for further analysis.
BEA (Backwall Echo Attenuator) Option allows independent gain control of the region under Gate B for backwall echo monitoring.
19ヤ Rack Mount Model.
Wide range of applications
A 6 kHz pulse repetition frequency, real-time analog and TTL outputs makes the USN 60 flaw detectors series ideal for a wide range of automated systems testing applications. The exclusive SmartView feature displays even the shortest echoes for critical scanning and rotating part inspections.
The quality, durability, dependability and ease of use that you have come to expect of Krautkramerメs popular USN Series of instruments remains. From rugged field inspections to high resolution thin measurements, long acoustically clean materials, and immersion systems, the USN 60 flaw detector family extends the range of applications that a portable instrument can perform. Furthermore the selectable 450V Square Wave Pulser satisfies a wide range of tough-to-penetrate applications, such as difficult to penetrate metallic applications and especially non-metals inspection like composite materials.
Tools for easy weld inspection
Color Leg allows easy identification of leg and skip distances for weld inspection
GRID mode dynamically changes bands of display backgroundcolors for each leg.
A-SCAN mode dynamically changes the color for each leg of the モliveヤ A-Scan
Weld Rating Calculation simplifies the rating of weld indications according to AWS Specification D1.1. (Formula D = A ヨ B ヨ C).
Trigonometric flaw location function with curvature correction automatically calculates depth, surface distance, and sound path to flaw along with the leg of the inspection when using angle beam probes. All TOF measurements can be displayed in mm, inches or µs.
SmartView function along with variable persistence freeze modes displays the most important information (relevant shot) for a test.
Real time (single shot) analog and TTL outputs handle a wide range of automated systems applications.
Choose from Four Freeze Modes: ALL, Peak Std, Compare or Envelope for optimum waveform evaluation and comparison.
Three Variable Persistence Modes are selectable in Freeze Envelope to visually assist flaw detection & evaluation for scanning and moving part inspections.
Compare frozen reference wave-forms to live A-Scans in different colors to easily interpret test results.
Documentation and recording
Store & preview a minimum of 200 user-named data sets with A-Scans for quick recall and instrument setup.
UltraDOC 4 software program for bi-directional communication with a PC for easy storage of data sets with A-scan and documentation of test results.
UltraMATEル software program simplifies the transfer, storage, analysis, and documentation of thickness data.
Reports with A-Scans are output directly to a variety of printers.
Specifications:
Range USN 60 0.040ヤ to 1100ヤ (1 mm to 28 m) at steel velocity; range selectable in fixed steps or continuously variable
Material Velocity Continuously adjustable from 0.0098 to 0.6299 inches/µs (250 to 16,000 m/s); 65 selectable material velocities
Display Delay -20 to 3498 µs in steel (dependent on range)
Probe Delay/Zero Offset 0 to 999.9 µs
Damping 50, 75, 150, 500 ohms
Gain 0 to 110 dB adjustable in selectable steps 0.1, 0.5, 1.0, 2.0, 6.0, user definable, and locked
Test Modes Pulse echo, dual, and thru-transmission
Pulser Square wave excitation pulse
Pulse Voltage (Square wave pulser mode) 50 to 450 V scrollable in 10 V adjustments
Pulse Width (Square wave pulser mode) Tunable from 50 to 1000 ns in 10 ns adjustments
Pulse Energy (Spike mode) Low, High
Pulse Repetition Frequency USN 60 Autolow, autohigh, manually adjustable from 15 to 6000 Hz (spike mode) and 15 to 2000 Hz in square wave mode, in 5 Hz increments, external trigger (spike mode only)
Bandwidth (amplifier bandpass) 0.25 to 25 MHz with 10 selectable settings including broadband
Gate Monitors Two independent flaw gates controllable over entire sweep range
Measurement Modes Zero-to-first, multi-echo with selectable flank or peak detection
Rectification Positive halfwave, negative halfwave, fullwave, RF
Reject (suppression) 0 to 80% linear
Units Inch, millimeter, or microsecond selectable
Operating Temperature -20° to 55°C (-4° to 130°F);
-25° to 70ºC (-13° to 158 º F ) storable
Languages Selectable English, German, French, Spanish, Italian, Portuguese, Norwegian, Swedish, Finnish, Danish, Dutch, Russian, Czech, Romanian, Slovakian
Probe Connectors BNC or Lemo
Keypad International symbols
Battery Power Lithium Ion battery pack; NiMH, NiCad or alkaline cells substitutable
Battery Life 11 hours on Li-Ion battery pack
Size 11.1ヤ W x 6.75ヤ H x 6.25ヤ D (282 x 171 x 159 mm)
Display 640 x 480 pixels Color LCD 132.48 x 99.36 mm
Weight 6.6 lbs. (3.0 kg) Li-Ion battery;
3.5 lbs.(1.6 kg) without battery
Color Leg Easy identification of leg and skip distances for angle beam inspection in A-scan or grid background colors
Weld Rating Calculation Simplifies the rating of weld indications according to AWS specification D1.1, (formula D=A-B-C)
Dust Proof/DrippingWater Proof As per IEC 529 specification for IP54 classification
Compliance EMC/EMI: EN 55011:2007, EN 61000-6-2:2005 Ultrasound: EN 12668, ASTM E317